K & Us Equipment logo  

Your source for quality
used semiconductor equipment

 


 
:: Home    :: Contact     :: About
K & Us Equipment, Inc., 30 Janis Way, Ste F, Scotts Valley, CA 95066 ¦ Phone (831) 461-9230, Fax (831) 461-9236 ¦ Contact www.kandus.com
 

Parker Balston 78-40 TOC Gas Generator

Parker Balston 78-40 TOC Gas Generator
Parker Balston 78-40 TOC Gas Generator
Parker Balston 78-40 TOC Gas Generator
Parker Balston 78-40 TOC Gas Generator
Parker Balston 78-40 TOC Gas Generator
Parker Balston 78-40 TOC Gas Generator

Parker Balston 78-40 TOC Gas Generator
Description: The Parker Balston 78-40 TOC Gas Generator produces carrier/combustion gas from an existing compressed air supply for TOC instruments, eliminating the need to purchase expensive, inconvenient, high pressure cylinders of oxygen, air or nitrogen.

The Parker Balston 78-40 utilizes catalytic oxidation and pressure swing adsorption technologies to remove hydrocarbons to 0.1 ppm (measured as methane), CO2 to 1 ppm and water vapour to 1 ppm (-73°C dewpoint).

The Parker Balston TOC Gas Generator eliminates all the inconvenience and cost of cylinder gas supplies and dependence on outside vendors. Uncontrollable vendor price increases, contract negotiations, long term commitments and cylinder rentals are no longer a concern. The Parker Balston TOC Gas Generator offers long term cost stability.

The 78-40 is a complete system with carefully matched components engineered for easy installation, operation and long term reliability. Installation consists of connecting a standard  compressed air line to the inlet and connecting the outlet to the TOC gas supply line. Plug the generator into a standard electrical wall outlet and within minutes high purity carrier/combustion gas is supplied.

Features:
Replaces high pressure oxygen or nitrogen gas cylinders with hydrocarbon-free, CO2-free compressed gas for TOC analysers.
Ensures consistent, reliable, TOC operation and significantly reduces instrument service and maintenance costs.
Compact design frees up valuable laboratory floor space.
Purity meets or exceeds all TOC manufacturers' gas purity requirements.

Principal Specifications:
Maximum TOC Gas Flow Rate (outlet) at 7barg:        1.25lpm
Outlet Hydrocarbon Concentration (as methane):        0.1ppm
Outlet CO2 concentration:                < 1ppm
Dewpoint:                        < -73C
Inlet and Outlet Port Connections:            1/4" NPT (female)
Minimum/Maximum inlet Air Pressure:            4.5 barg/8.6 barg
Maximum Inlet Air Temperature:                25C
Minimum Required Inlet Air Flow at 7 barg:        2.5 lpm (2500ml/min)
Maximum Inlet Hydrocarbon Concentration (as methane):    100 ppm
Pressure Drop at Maximum Flow Rate:            0.5 barg
Warm-up Time:                        40 minutes
Electrical Requirements:                 115V/60Hz
Dimensions:                         12"w X 17.5"d X 16.6"h


Price: Inquire
Ref# 37624ke

Request a Quote     Contact Us

Back

K & Us Equipment, Inc., 30 Janis Way, Ste F, Scotts Valley, CA 95066 ¦ Phone (831) 461-9230, Fax (831) 461-9236 ¦ Contact www.kandus.com