Description: Met One's One CFM Particle Counters are reliable and easy to use.

The A2408 has 0.3 micron sensitivity and 2, 4, 5, or 6 channels. The uilt-in printer (DOES NOT WORK-SEE BELOW) will print as you work and you can also "smart" print stored count data as counts-per-cubic foot, total counts, or Fed-Std 209E calculations. Data can be stored with numerical location labels (0-999).
b
All operational features are accessed from a simple, well organized front panel.
The A2408 supports the Model 2431 Manifold which permits the A2408 to automatically sample up to 32 locations.

This Particle counter appears to work very well and is very clean. Please note that the built-in printer doesn't appear to work. However, there are instructions in the manual on how to hook up an external printer via the RS-232 port.

Also, there is software available at: http://www.hachco.ca/met-one-portall-software/product?id=14534105393

MET ONE PortAll Software

Working with your handheld and portable device, PortAll Version 2 simply and easily organizes archives and retrieves particle count data. Intuitive Windows-based software dramatically reduces operator training time. Track particle count data, identify trends and eliminate particle contamination sources.

  • Schedule and automate handheld and portable devices
  • Gown up less
  • Report with ease
  • Simplify validation
  • Expand your particle counter’s capabilities

Specifications:
Model A2408-1-115-1
Size Channels:
    0.3 Micron:   
        2 ch:    0.3, 0.5
        4 ch:    0.3, 0.5, 1.0, 5.0
        5 ch:    0.3, 0.5, 1.0, 5.0, 10.0
        6 ch:    0.3, 0.5, 1.0, 3.0, 5.0, 10.0
  
Light Source:    Laser diode (30,000 hours mean time between failures)
Outputs:    RS232, RS485 serial interface
Sample Time:    1 second to 24 hours
Alarm Counts:    1-9,999,999
Location
   Labels:    0-999 appear on printout
Data Storage:    400 Samples
Size:        11.2" W x 6.0"H x 18" D
Weight:        24 pounds
Environment:
  Operating:    55 to 105°F (12 to 41°C)
        29 to 95% relative humidity, noncondensing
  Storage:    -40 to 160°F (-40 to 70°C)
        up to 98% relative humidity, noncondensing
Power:        115VAC  1A  50/60Hz